Chiller for microscope

Double level of precision and stability

 

Microscopy universe needs a high precision temperature control. To meet the cooling needs of the transmission electron microscopy, Eurodifroid designed a cooling solution consisting two precision levels

A first cooler located outside the building, with a frost kit and a control level of +/- 0.1 ° C, which supply 4 end line coolers themselves benefits a level of precision of +/- 0.1 ° C, this end line of each cool a TEM microscope.

EURODIFROID solution

*Datas given at 20°C water outlet, 32°C ambient temperature
**Datas given at 20°C water outlet, 10°C water inlet in primary circuit
Product RFC250 + ECH45
Quantity 1 RFC pour 4 ECH
Cooling capacity RFC 25000 W* + ECH 4500 W**
Technology PID water chiller + End line chiller

Diagramm

Refroidisseur pour microscope

Technical datas


*Datas given at 10°C water outler, 32°C ambient temperature
RFC250 THBINVWEXTT
Cooling capacity* 25 000 W
Supply power 400-3-50Hz
Running current 31,1 A
Starting current 50 A
Line protection 40 A
Power consumption 16868 W
Ambient temperature mini/maxi -15/40°C
Control temperature mini/maxi 10/25°C
Pump flow mini/maxi 40-160 l/min
Pump pressure mini/maxi 1,8-4 bar
Tank capacity 100 l
Refrigerant R410a
Noise <70dB(A)
Weight 365 kg
ECH45 AWBIPIDGAR
**Datas given at 20°C water outlet, 10°C water inlet
Cooling capacity** 4 500 W
Supply power 230-1-50Hz
Running current 2,4 A
Starting current 10 A
Line protection 4 A
Power consumption 470 W
Ambient temperature mini/maxi 10/40°C
Flow mini primary circuit 10 l/min
flow mini/maxi secondary circuit 3-41 l/min
Pressure maxi primary circuit 5 bar
Tank capacity 30 l
Refrigerant primary/secondary circuit Eau
Noise <70dB(A)
Weight 50 kg

Temperature stability example

Water chiller for mircoscope